SOP for Temperature & Humidity Monitoring in Quality Control (QC) Department
🌡️ SOP: Temperature & Humidity Monitoring in QC Department
🎯 Purpose
This SOP provides a standardized and regulated method for monitoring, recording, reporting, and controlling environmental conditions (temperature and humidity) in QC areas to ensure sample integrity, instrument performance, and consistent laboratory operations.
📌 Scope
This procedure applies to all controlled QC areas where ambient conditions may impact:
- Testing outcomes
- Sample stability
- Instrument performance
🔤 Abbreviations
- QC: Quality Control
- QA: Quality Assurance
- RH: Relative Humidity
- HVAC: Heating, Ventilation & Air Conditioning
📝 Definitions
- Temperature: Degree of hotness or coldness in controlled environments
- Relative Humidity (RH): % of moisture in air relative to saturated steam at the same temperature
- Hygrometer: Calibrated digital device for continuous monitoring of temperature and RH
👥 Responsibilities
| Role | Responsibility |
|---|---|
| QC Analyst/Executive | Ensure hygrometer remains ON continuously and record temperature & RH in log sheets |
| QC Supervisor | Review daily logs, ensure completeness, and initiate corrective actions for excursions |
| Quality Assurance (QA) | Review trends, deviations, and ensure compliance |
| Engineering / Maintenance | Ensure HVAC systems and monitoring devices are functional and calibrated |
⚙️ Equipment & Materials
- Calibrated Digital Hygrometer / Thermo-Hygrometer
- Authorized Temperature & Humidity Log Sheet
- Valid Calibration Certificates
- Backup power supply (if applicable)
📝 Procedure
7.1 Preparation
- Confirm hygrometer is calibrated and installed at a representative location
- Switch ON and allow device to stabilize before taking readings
7.2 Monitoring Frequency
- Continuous monitoring using data logger (preferred)
- Manual recording: At least twice daily — start of shift and mid-shift/closing
7.3 Recording Readings
- Record on authorized log sheet:
- Date
- Time
- Temperature (°C)
- Relative Humidity (RH %)
- Recorder’s Name/Signature
✔ Acceptance Criteria
| Parameter | Acceptable Range |
|---|---|
| Temperature | 20°C – 25°C |
| Relative Humidity (RH) | 40% – 60% RH |
Note: Site-specific authorized limits may vary and should be justified within the log sheet.
⚠️ Deviations & Corrective Actions
- If temperature or RH values fall outside limits, notify QC Supervisor & Engineering immediately
- Investigate cause (e.g., HVAC failure, doors/windows open)
- Take corrective action (adjust HVAC, relocate samples, etc.)
- Document deviation, cause, action taken, and impact assessment in deviation report
🔧 Calibration & Maintenance
- Calibrate hygrometers as per schedule (typically every 6–12 months)
- Maintain valid calibration certificates
- Check device functionality and battery/back-up power regularly
📁 Records & Retention
- Original Temperature & Humidity Logs
- Calibration Certificates
- Deviation Reports
- Trend Analysis Reports
- All records maintained per Good Documentation Practices (GDP) and audit-ready
🎓 Training
- Personnel performing monitoring must be trained on this SOP
- Training records must be maintained and periodically reviewed
- Refresher training conducted after updates or deviations
📎 Annexure
- Annexure-I: Temperature & Humidity Monitoring Log Sheet
Comments
Post a Comment