SOP for Temperature & Humidity Monitoring in Quality Control (QC) Department

🌡️ SOP: Temperature & Humidity Monitoring in QC Department


🎯 Purpose

This SOP provides a standardized and regulated method for monitoring, recording, reporting, and controlling environmental conditions (temperature and humidity) in QC areas to ensure sample integrity, instrument performance, and consistent laboratory operations.

📌 Scope

This procedure applies to all controlled QC areas where ambient conditions may impact:

  • Testing outcomes
  • Sample stability
  • Instrument performance

🔤 Abbreviations

  • QC: Quality Control
  • QA: Quality Assurance
  • RH: Relative Humidity
  • HVAC: Heating, Ventilation & Air Conditioning

📝 Definitions

  • Temperature: Degree of hotness or coldness in controlled environments
  • Relative Humidity (RH): % of moisture in air relative to saturated steam at the same temperature
  • Hygrometer: Calibrated digital device for continuous monitoring of temperature and RH

👥 Responsibilities

Role Responsibility
QC Analyst/Executive Ensure hygrometer remains ON continuously and record temperature & RH in log sheets
QC Supervisor Review daily logs, ensure completeness, and initiate corrective actions for excursions
Quality Assurance (QA) Review trends, deviations, and ensure compliance
Engineering / Maintenance Ensure HVAC systems and monitoring devices are functional and calibrated

⚙️ Equipment & Materials

  • Calibrated Digital Hygrometer / Thermo-Hygrometer
  • Authorized Temperature & Humidity Log Sheet
  • Valid Calibration Certificates
  • Backup power supply (if applicable)

📝 Procedure

7.1 Preparation

  • Confirm hygrometer is calibrated and installed at a representative location
  • Switch ON and allow device to stabilize before taking readings

7.2 Monitoring Frequency

  • Continuous monitoring using data logger (preferred)
  • Manual recording: At least twice daily — start of shift and mid-shift/closing

7.3 Recording Readings

  • Record on authorized log sheet:
    • Date
    • Time
    • Temperature (°C)
    • Relative Humidity (RH %)
    • Recorder’s Name/Signature

✔ Acceptance Criteria

Parameter Acceptable Range
Temperature 20°C – 25°C
Relative Humidity (RH) 40% – 60% RH

Note: Site-specific authorized limits may vary and should be justified within the log sheet.

⚠️ Deviations & Corrective Actions

  • If temperature or RH values fall outside limits, notify QC Supervisor & Engineering immediately
  • Investigate cause (e.g., HVAC failure, doors/windows open)
  • Take corrective action (adjust HVAC, relocate samples, etc.)
  • Document deviation, cause, action taken, and impact assessment in deviation report

🔧 Calibration & Maintenance

  • Calibrate hygrometers as per schedule (typically every 6–12 months)
  • Maintain valid calibration certificates
  • Check device functionality and battery/back-up power regularly

📁 Records & Retention

  • Original Temperature & Humidity Logs
  • Calibration Certificates
  • Deviation Reports
  • Trend Analysis Reports
  • All records maintained per Good Documentation Practices (GDP) and audit-ready

🎓 Training

  • Personnel performing monitoring must be trained on this SOP
  • Training records must be maintained and periodically reviewed
  • Refresher training conducted after updates or deviations

📎 Annexure

  • Annexure-I: Temperature & Humidity Monitoring Log Sheet

Comments